Accuracy Analysis and Error Correction for Anatomical Landmarks Based Registration in Image-Guided Neurosurgery.pdf_第1頁(yè)
已閱讀1頁(yè),還剩95頁(yè)未讀 繼續(xù)免費(fèi)閱讀

下載本文檔

版權(quán)說(shuō)明:本文檔由用戶提供并上傳,收益歸屬內(nèi)容提供方,若內(nèi)容存在侵權(quán),請(qǐng)進(jìn)行舉報(bào)或認(rèn)領(lǐng)

文檔簡(jiǎn)介

1、Accurate patient to image registration is the core for successful and safe image-guided neuro-navigation.Point-matching is the most common technique in practice to achieve this registration.While SkinAdhesive Markers (SM

2、s) are widely used in point-matching registration, a proper implementation ofAnatomical Landmarks (ALs) may overcome the inconvenience brought by the use of SMs.However,the accuracy and applicability of ALs registration

3、in neuro-navigation have remained a controversialissue due to the great variability of the reported results and conclusions.Therefore, we are providing ananalysis for the accuracy when using different sets of ALs as well

4、 as the applicability of thecorresponding surgical fields.In addition, one of the significant factors that influence the registrationaccuracy at the target point is the distribution of the fiducial points.The optimal dis

5、tribution may bedifficult to achieve either due to the limited number of distinct anatomical features on head surface oreven due to the poor planning of skin adhesive markers.Therefore we have also tried to overcome this

6、problem and correct the quality of the registration in the areas that normally suffer low accuracy whenusing ALs-based registration.
  In the accuracy analysis study, we propose a set of three configurations using nin

7、e ALs.Theseconfigurations are defined according to the required positioning of the patient's head during surgery andthe resulting distribution of the expected Target Registration Error (TRE).We first evaluated theseconfi

8、gurations by simulation experiments using real clinical data of 20 patients from two hospitals, andthen tested the applicability of them in eight real clinical surgeries of neuronavigation.
  In the correction study,

9、the proposed method is based on an initial point-matching registration using sixdistinct ALs as fiducial points, followed by selection of some surface points on the patient's head atlocations where natural ALs are not av

10、ailable to improve the distribution of the fiducial points.Theprojection of the surface points from patient space into image space reflects the errors introduced duringthe initial registration process.These errors can be

11、 identified and inspected in the image space bycalculating the distance between the projected surface point positions transferred with the standardspatial relation and also by calculating their nearest point positions on

12、 the head surface.This informationis then used to improve registration accuracy by adopting the calculated nearest points (the moreaccurate points) in the image space, instead of the projected positions, along with their

13、 correspondingactual surface points in the patient space, as additional pairs of registration points.
  The simulation experiment of the accuracy analysis showed that, by incorporating a FiducialRegistration Error (FR

14、E) of 3.5 mm measured in the clinical setting, the expected TRE in the wholeskull was less than 2.5 mm, and the expected TRE in the whole brain was less than 1.75 mm when usingthe configuration with all the nine ALs.A sm

15、all TRE could also be achieved in the correspondingsurgical field, by using the other three configurations with less ALs.In the clinical experiment, theFiducial Localization Error (FLE) ranges in the image and the patien

16、t space were 1.4-3.6 mm and 1.6-5.5 mm, respectively.The measured TRE and FRE were 3.1 ± 0.75 mm and 3.5 ± 0.17 mm, respectively.
  For the correction trials, experiments with real clinical data showed that when using

17、 the surface points tocorrect the initial registration transformation, the TRE decreased in the whole-brain area byapproximately 20%, and this improvement is more dominant in the posterior and superior parts of thebrain.

18、
  The ALs configurations proposed in the accuracy analysis provide sufficient registration accuracy andcan help to avoid SMs disadvantages if used clinically.The correction technique helps as well toovercome the natu

19、rally impaired distribution of the ALS, which is the most significant factor thatprevents their wide use in registration.The method also allows more precise selection of correspondingfiducial points than traditional ALs

溫馨提示

  • 1. 本站所有資源如無(wú)特殊說(shuō)明,都需要本地電腦安裝OFFICE2007和PDF閱讀器。圖紙軟件為CAD,CAXA,PROE,UG,SolidWorks等.壓縮文件請(qǐng)下載最新的WinRAR軟件解壓。
  • 2. 本站的文檔不包含任何第三方提供的附件圖紙等,如果需要附件,請(qǐng)聯(lián)系上傳者。文件的所有權(quán)益歸上傳用戶所有。
  • 3. 本站RAR壓縮包中若帶圖紙,網(wǎng)頁(yè)內(nèi)容里面會(huì)有圖紙預(yù)覽,若沒(méi)有圖紙預(yù)覽就沒(méi)有圖紙。
  • 4. 未經(jīng)權(quán)益所有人同意不得將文件中的內(nèi)容挪作商業(yè)或盈利用途。
  • 5. 眾賞文庫(kù)僅提供信息存儲(chǔ)空間,僅對(duì)用戶上傳內(nèi)容的表現(xiàn)方式做保護(hù)處理,對(duì)用戶上傳分享的文檔內(nèi)容本身不做任何修改或編輯,并不能對(duì)任何下載內(nèi)容負(fù)責(zé)。
  • 6. 下載文件中如有侵權(quán)或不適當(dāng)內(nèi)容,請(qǐng)與我們聯(lián)系,我們立即糾正。
  • 7. 本站不保證下載資源的準(zhǔn)確性、安全性和完整性, 同時(shí)也不承擔(dān)用戶因使用這些下載資源對(duì)自己和他人造成任何形式的傷害或損失。

評(píng)論

0/150

提交評(píng)論